Ultra-high-b-value diffusion-weighted MR imaging for the detection of prostate cancer: evaluation in 201 cases with histopathological correlation

Eur Radiol. 2011 Jan;21(1):188-96. doi: 10.1007/s00330-010-1883-7. Epub 2010 Jul 18.

Abstract

Purpose: To assess the diagnostic performance of diffusion-weighted magnetic resonance (MR) imaging (DWI) for prostate cancer detection, using different b-values.

Methods: A total of 201 patients who underwent MR imaging before total prostatectomy were evaluated. MR images were independently assessed by three radiologists. Three combinations of sequences were separately evaluated, as follows: group 1 [T2-weighted images (T2WI) alone], group 2 (T2WI and DWI with a b-value of 1,000 s/mm2), group 3 (T2WI and DWI with a b-value of 2,000 s/mm2). Whole-mount-section histopathological examination was the reference standard. Areas under the receiver operating characteristic curve (AUCs) and diagnostic performance parameters were determined.

Results: The sensitivity, specificity, and AUC for the detection of prostate cancer were as follows: 52.2%, 80.7%, and 0.694 in group 1; 61.2%, 82.6%, and 0.755 in group 2; 73.2%, 89.7%, and 0.842 in group 3. Group 3 achieved the highest diagnostic performance, followed by group 2 (P<0.05). In the transition zone, the specificity was lower (P<0.001) for group 2 (82.2%) than for group 1 (86.2%).

Conclusion: The addition of diffusion-weighted images with a b-value of 2,000 s/mm2 to T2WI can improve the diagnostic performance of MR imaging in prostate cancer detection.

MeSH terms

  • Adult
  • Aged
  • Diffusion Magnetic Resonance Imaging*
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Prostatic Neoplasms / diagnosis*
  • Prostatic Neoplasms / diagnostic imaging
  • Prostatic Neoplasms / pathology*
  • Radiography