Purpose: To assess the stability of the Akreos AO intraocular lens (IOL) platform with a simulated toric design using objective image analysis.
Setting: Six hospital eye clinics across Europe.
Methods: After implantation in 1 eye of patients, IOLs with orientation marks were imaged at 1 to 2 days, 7 to 14 days, 30 to 60 days, and 120 to 180 days. The axis of rotation and IOL centration were objectively assessed using validated image analysis.
Results: The study enrolled 107 patients with a mean age of 69.9 years +/- 7.7 (SD). The image quality was sufficient for IOL rotation analysis in 91% of eyes. The mean rotation between the first day postoperatively and 120 to 180 days was 1.93 +/- 2.33 degrees, with 96% of IOLs rotating fewer than 5 degrees and 99% rotating fewer than 10 degrees. There was no significant rotation between visits and no clear bias in the direction of rotation. In 71% of eyes, the dilation and image quality was sufficient for image analysis of centration. The mean change in centration between 1 day and 120 to 180 days was 0.21 +/- 0.11 mm, with all IOLs decentering less than 0.5 mm. There was no significant decentration between visits and no clear bias in the direction of the decentration.
Conclusion: Objective analysis of digital retroillumination images taken at different postoperative periods shows the aspheric IOL platform was stable in the eye and is therefore suitable for the application of a toric surface to correct corneal astigmatism.