Hydroxyapatite-coated pins versus titanium alloy pins in external fixation at the wrist: a controlled cohort study

J Trauma. 2011 Apr;70(4):845-51. doi: 10.1097/TA.0b013e3181e97761.

Abstract

Background: The purpose of this study was to analyze whether the prevalence of pin-related complications can be reduced by the use of hydroxyapatite (HA)-coated pins in external fixators applied for unstable wrist fractures.

Methods: Forty patients (160 pins) were randomized for standard uniplanar fixator treatment with the use of identically designed pins either composed of titanium-alloy (Ti6Al4V) (n = 20) or coated by HA (n = 20). Each pin site was clinically evaluated with regard to erythema, drainage, pain value, and radiologically assessed concerning loosening at T1 (mean, 9 days), T2 (mean, 43 days), and T3 (mean, 56 days). In case of pin-track complication, the patient was followed continuously. The need for antibiotics or additional surgery was documented. Bone mineral density was analyzed by Dual Energy X-ray Absorptiometry. At fixator removal (T2), the pin-extraction strength was measured by the use of a digital-torque-wrench.

Results: Two minor pin-track infections requiring oral antibiotics occurred in the HA-pin group (2.7%) (p > 0.05). The vast majority of clinical pin-site parameters were comparable in both groups. At the end of the fixator therapy, there were 16 loose pins (n(Ti6AL4V-group) = 10; n(HA-group) = 6). The rate of loose pins was correlated to patient's age (p < 0.05) but not to bone mineral density values or the occurrence of pin-site infection. Finally, no significant difference between the two groups was detected with regard to the prevalence of clinical relevant pin-site complications (p = 0.80).

Conclusions: In external fixation of the wrist, the use of HA-coated pins yields no clinical advantages: there is a trend toward a superior pin-bone anchorage, but a tendency of increased susceptibility for minor pin-track infections.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Alloys*
  • Bone Nails*
  • Coated Materials, Biocompatible*
  • Durapatite*
  • Female
  • Fracture Fixation / instrumentation*
  • Fractures, Bone / diagnosis
  • Fractures, Bone / surgery*
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Treatment Outcome
  • Wrist Injuries / diagnosis
  • Wrist Injuries / surgery*
  • Young Adult

Substances

  • Alloys
  • Coated Materials, Biocompatible
  • titanium aluminum alloy
  • Durapatite