Electronic working length determination in primary teeth by ProPex and Digital Signal Processing

Aust Endod J. 2010 Dec;36(3):105-8. doi: 10.1111/j.1747-4477.2009.00185.x. Epub 2010 Aug 23.

Abstract

The purpose of this study was to evaluate the accuracy of electronic apex locators Digital Signal Processing (DSP) and ProPex, for root canal length determination in primary teeth. Fifteen primary molars (a total of 34 root canals) were divided into two groups: Group I - without physiological resorption (n = 16); and Group II - with physiological resorption (n = 18). The length of each canal was measured by introducing a file until its tip was visible and then it was retracted 1 mm. For electronic measurement, the devices were set to 1 mm short of the apical resorption. The data were analysed statistically using the intraclass correlation coefficient (ICC). Results showed that the ICC was high for both electronic apex locators in all situations - with (ICC: DSP = 0.82 and Propex = 0.89) or without resorption (ICC: DSP = 0.92 and Propex = 0.90). Both apex locators were extremely accurate in determining the working length in primary teeth, both with or without physiological resorption.

Publication types

  • Comparative Study

MeSH terms

  • Dental Pulp Cavity / anatomy & histology*
  • Humans
  • Molar / anatomy & histology
  • Odontometry / instrumentation*
  • Odontometry / methods
  • Root Resorption
  • Signal Processing, Computer-Assisted*
  • Tooth Root / anatomy & histology*
  • Tooth, Deciduous / anatomy & histology*