Trimming of a Migrated Biliary Nitinol Stent Using Argon Plasma

Case Rep Gastroenterol. 2009 Jul 31;3(2):202-206. doi: 10.1159/000226252.

Abstract

Metallic stent migration is a well-known complication which cannot always be managed by removal or repositioning, especially in case of uncovered stent. We report a patient who developed obstructive jaundice due to migration of an expandable metallic stent (EMS) inserted in the lower bile duct. Trimming of the EMS using argon plasma was performed, with the power setting of 60 W and 2.0 l/min of argon flow. The distal part of the EMS was removed and mechanical cleaning using balloon catheter was performed for remnant EMS. Without additional stent insertion, jaundice was relieved in a few days. No complication was recognized during the procedure and no recurrence of jaundice in the rest of his life.

Publication types

  • Case Reports