Salivary testosterone levels in Korean adults were quantitatively measured for the first time by liquid chromatography-electrospray-tandem mass spectrometry (LC ESI MS/MS). Salivary testosterone was separated on a multiple reaction monitoring (MRM) chromatogram within 7 min. The LC ESI MS/MS assay was validated over the linearity range of 0.01-2.00 ng/ml (r=0.99987) using testosterone-d(3) as an internal standard. The lower limit of quantification (LOQ) was 0.01 ng/ml. The intra- and inter-assay precisions were 1.54% to 4.09% and 0.96% to 4.29%, respectively. The mean recovery was 93.32% (range 88.43-98.05%). The validated assay was then applied to measure the salivary testosterone levels of Korean adults. In men, the salivary testosterone level collected between 9:00-11:00 am was approximately 2.8 times higher than that in women (P < 0.0001). Salivary testosterone levels in both sexes negatively correlated with age. The present assay would also be useful in measuring salivary testosterone levels in clinical laboratories.