Calibration of an in-situ BEGe detector using semi-empirical and Monte Carlo techniques

Appl Radiat Isot. 2011 Aug;69(8):1151-5. doi: 10.1016/j.apradiso.2010.12.005. Epub 2010 Dec 16.

Abstract

In the case of a nuclear or radiological accident a rapid estimation of the qualitative and quantitative characteristics of the potential radioactive pollution is needed. For aerial releases the radioactive pollutants are finally deposited on the ground forming a surface source. In this case, in-situ γ-ray spectrometry is a powerful tool for the determination of ground pollution. In this work, the procedure followed at the Nuclear Engineering Department of the National Technical University of Athens (NED-NTUA) for the calibration of an in-situ Broad Energy Germanium (BEGe) detector, for the determination of gamma-emitting radionuclides deposited on the ground surface, is presented. BEGe detectors due to their technical characteristics are suitable for the analysis of photons in a wide energy region. Two different techniques were applied for the full-energy peak efficiency calibration of the BEGe detector in the energy region 60-1600 keV: Full-energy peak efficiencies determined using the two methods agree within statistical uncertainties.