Feasibility of in vivo measurement of 239 Pu distribution in lungs using an imaging plate

Appl Radiat Isot. 2011 May;69(5):808-13. doi: 10.1016/j.apradiso.2011.01.018. Epub 2011 Jan 21.

Abstract

An in vivo measurement system using an imaging plate (IP) system was developed, which displayed images reflecting (239)Pu distribution in the lung of a phantom. The detection limits of the IP system for 1-12h exposures were between 1670 and 245 Bq at a 1.6 cm chest wall thickness. The detection limit of the IP system for a 2.5h exposure was equal to that of a germanium detector for a 0.5h measurement. The IP system could be used as a new device for in vivo measurement of (239)Pu in the lung.

MeSH terms

  • Equipment Design
  • Feasibility Studies
  • Limit of Detection
  • Lung / chemistry*
  • Plutonium / analysis*
  • Time Factors