Unmappable ventricular tachycardia after an old myocardial infarction. Long-term results of substrate modification in patients with an implantable cardioverter defibrillator

J Interv Card Electrophysiol. 2011 Aug;31(2):149-56. doi: 10.1007/s10840-011-9549-1. Epub 2011 Feb 22.

Abstract

Purpose: The frequent occurrence of ventricular tachycardia can create a serious problem in patients with an implantable cardioverter defibrillator. We assessed the long-term efficacy of catheter-based substrate modification using the voltage mapping technique of infarct-related ventricular tachycardia and recurrent device therapy.

Methods: The study population consisted of 27 consecutive patients (age 68 ± 8 years, 25 men, mean left ventricular ejection fraction 31 ± 9%) with an old myocardial infarction and multiple and/or hemodynamically not tolerated ventricular tachycardia necessitating repeated device therapy. A total of 31 substrate modification procedures were performed using the three-dimensional electroanatomical mapping system. Patients were followed up for a median of 23.5 (interquartile range 6.5-53.2) months before and 37.8 (interquartile range 11.7-71.8) months after ablation. Antiarrhythmic drugs were not changed after the procedure, and were stopped 6 to 9 months after the procedure in patients who did not show ventricular tachycardia recurrence.

Results: Median ventricular tachycardias were 1.6 (interquartile range 0.7-6.7) per month before and 0.2 (interquartile range 0.00-1.3) per month after ablation (P = 0.006). Nine ventricular fibrillation episodes were registered in seven patients before and two after ablation (P = 0.025). Median antitachycardia pacing decreased from 1.6 (interquartile range 0.01-5.5) per month before to 0.18 (interquartile range 0.00-1.6) per month after ablation (P = 0.069). Median number of shocks decreased from 0.19 (interquartile range 0.04-0.81) per month before to 0.00 (interquartile range 0.00-0.09) per month after ablation (P = 0.001). One patient had a transient ischemic attack during the procedure, and another developed pericarditis. Nine patients died during follow-up, eight patients due to heart failure and one patient during valve surgery.

Conclusion: Catheter-based substrate modification using voltage mapping results in a long-lasting reduction of cardioverter defibrillator therapy in patients with multiple and/or hemodynamically not tolerated infarct-related ventricular tachyarrhythmia.

MeSH terms

  • Academic Medical Centers
  • Aged
  • Body Surface Potential Mapping / methods
  • Catheter Ablation / methods*
  • Cohort Studies
  • Combined Modality Therapy
  • Defibrillators, Implantable*
  • Electrocardiography
  • Female
  • Follow-Up Studies
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Myocardial Infarction / complications*
  • Myocardial Infarction / diagnosis
  • Recurrence
  • Retrospective Studies
  • Risk Assessment
  • Statistics, Nonparametric
  • Tachycardia, Ventricular / diagnosis*
  • Tachycardia, Ventricular / etiology
  • Tachycardia, Ventricular / therapy*
  • Time Factors
  • Treatment Outcome