Background and aim of the study: Transcatheter aortic valve implantation (TAVI) has been introduced for the treatment of severe symptomatic aortic stenosis in patients not suitable for surgical valve replacement. However, a potential problem of TAVI is the development of severe aortic insufficiency after valve implantation due to a too-low implantation of the valve.
Methods: Since August 2008, a total of 33 TAVI procedures using the 18 Fr Medtronic CoreValve ReValving system has been performed at the authors' institution.
Results: Severe post-implantation aortic regurgitation occurred in three patients (9%), due to a too-low implantation. Two of these patients underwent a catheter-based repositioning of the valve using a standard snare; the third patient declined any further intervention. Both repositioning procedures were uneventful, with no significant residual regurgitation.
Conclusion: Severe aortic regurgitation after TAVI with the Medtronic CoreValve system is not uncommon. If the valve is implanted too low, a catheter-based valve repositioning may be the method of choice to resolve the problem.