Surgical pitfalls encountered with the semi-implantable middle ear transducer ossicular stimulator (MET-Otologics)

Rev Laryngol Otol Rhinol (Bord). 2010;131(3):193-7.

Abstract

Objective: To determine whether surgical pitfalls can be anticipated through a preoperative CT-Scan in case of middle ear implantation with a semi-implantable middle ear ossicular stimulator from Otologics (SIMOS). A second outcome measure was to assess the auditory gain obtained in the series.

Materials and methods: A retrospective analysis of 19 consecutive surgical procedure for a SIMOS implantations was achieved. Preoperative Ct-Scans were analyzed and several measurements of the mastoid were taken out from axial transverse views, ie. the distance between the incus and the cortical mastoid bone [d(i-c)], the distance between the sigmoid sinus and the incus [d(i-ss)], the distance between the sigmoid sinus and the cortical bone of the mastoid [d(c'-ss)]. On coronal images, the dura of the middle fossa was described as procident if it was lower than the head of malleus and/or body incus. Auditory gain was calculated on pure tone and speech audiometry with bisyllabic words.

Results: All cases were successfully implanted. d(i-c) was on average at 33.1 +/- 2.6 mm in cases easy to implant. One case of contracted mastoid was difficult and led to the impaction of the sigmoid sinus. In this case the distance d(i-c) was shorter at 25 mm than the average minus 2 standard deviations of the others (27.9 mm). Also d(ss-c') was lower at 7 mm than the average minus 2 standard deviations of the others (7.6 mm). Two other cases led to difficult surgical procedures because of a low middle fossa dura. This too low middle fossa dura could be identified on preoperative coronal CT-scans in these two cases. Audiometric gains were remarkable, with an average of 39 +/- 16 dB. In case of severe sensorineural hearing loss the average gain was higher at 46 +/- 9 dB. Two revision-procedures were required, one for a device failure, the second because of skin alteration due to systemic inflammatory disease requiring corticosteroids. Mean follow-up was 50 +/- 22 months.

Conclusion: The SIMOS is a powerful middle ear implant that can undoubtedly give a huge audiometric gain. The implantation procedure requires a precise microsurgical operation that can be hampered by anatomical conditions of the petrous bone. A thorough examination of preoperative CT-scan is highly recommended in order to obtain measurements of the mastoid in axial views, and the analysis of the position of the middle fossa dura in coronal views.

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Audiometry, Pure-Tone
  • Audiometry, Speech
  • Cochlea
  • Cochlear Implantation
  • Cochlear Implants*
  • Ear, Middle
  • Female
  • Follow-Up Studies
  • Humans
  • Incus
  • Male
  • Malleus
  • Middle Aged
  • Preoperative Care
  • Retrospective Studies
  • Time Factors
  • Tomography, X-Ray Computed
  • Transducers*
  • Treatment Outcome