A convenient neutron source is made for calibration of neutron survey instruments and personal dosimeters that are used in various nuclear installations such as fuel reprocessing, waste management, fuel fabrication and oil and well logging facilities, etc. This source consists of a bare (241)Am-Be neutron source placed at the centre of a 15-cm radius stainless steel spherical shell filled with distilled water. This paper describes the standardisation of the source at Bhabha Atomic Research Centre, using De Pangher neutron long counter both experimentally and using the Monte Carlo simulation. The ratio of neutron yield of water moderated to the bare (241)Am-Be neutron source was found to be 0.573. From the simulation, the neutron-fluence-weighted average energy of water-moderated (241)Am-Be source (fluence-weighted average energy of 2.25 MeV, dose-weighted average energy of 3.55 MeV) was found to be nearly the same as that of a (252)Cf source (fluence-weighted average energy of 2.1 MeV, dose-weighted average energy of 2.3 MeV). This source can be used for calibration in addition to (252)Cf, to study the variation in response of neutron monitoring instruments.