Dislocation of a transapically implanted aortic valve prosthesis with a functionally bicuspid aortic valve and ascending aortic aneurysm

Thorac Cardiovasc Surg. 2011 Dec;59(8):503-6. doi: 10.1055/s-0030-1270995. Epub 2011 Apr 20.

Abstract

In recent years, catheter-based aortic valve interventions have become established procedures for the treatment of high-risk and advanced age patients with aortic valve pathologies. One of the limitations of the widespread applicability of this procedure is the annulus size. Until recently, no prosthesis was available to treat patients with a large annulus. We report on a patient with high-grade aortic stenosis (AS) and a 27-mm annulus, who underwent transapical implantation (TAP) of an Edwards SAPIEN® 29-mm prosthesis (Edwards LifeScience, Irvine, CA, USA). Due to insufficient dilation of his heavily calcified, functionally bicuspid aortic valve leaflets during balloon aortic valvuloplasty (BAV), the TAP prosthesis did not anchor adequately. This was determined during follow-up as he developed progressive aortic insufficiency and orthopnea, and an echocardiography revealed that the valve had been displaced into the LVOT. A conventional aortic valve replacement and ascending aorta replacement were performed, at which time the TAP prosthesis was removed. The patient recovered uneventfully, and was discharged with a well-functioning aortic bioprosthetic valve and in good general condition.

Publication types

  • Case Reports

MeSH terms

  • Aged, 80 and over
  • Aorta / surgery*
  • Aortic Aneurysm / surgery*
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / etiology
  • Aortic Valve Insufficiency / surgery*
  • Aortic Valve Stenosis / surgery*
  • Dyspnea / etiology
  • Follow-Up Studies
  • Heart Valve Prosthesis Implantation / adverse effects*
  • Humans
  • Male
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Treatment Outcome