Background/objectives: The purpose of this prospective, randomized, single-blind controlled clinical trial was to compare the effectiveness of a zotarolimus-eluting stent (ZoMaxx™) with a paclitaxel-eluting coronary stent (Taxus™ Express(2)™) in patients with angina pectoris and a single native coronary artery lesion between 10-28 mm in length and 2.5-3.75mm in diameter.
Methods: Patients were enrolled at 75 international institutions between June 2005 and November 2006.
Results: 1099 (1672 originally planned) patients received 557 ZoMaxx and 542 Taxus stents: cohorts were well-matched for diabetes (27% vs. 27%), reference vessel diameter (2.73 ± 0.46mm vs. 2.74 ± 0.45mm) and lesion length (14.8 ± 6.7mm vs. 14.3 ± 6.4mm). Nine month clinical and angiographic follow-up was available in 1052/1099 (96%) and 649/836 (78%) patients, respectively. The safety profiles for the two stents (myocardial infarction (MI), cardiac death and/or target vessel revascularization (TVR)) were similar (ZoMaxx 8.7% vs. Taxus 6.9%, p=NS). The primary endpoint of 9-month TVR occurred more frequently after treatment with ZoMaxx (6.8%) as compared with Taxus (4.2%), therefore the primary clinical endpoint was not met. However, the 9-month in-segment late lumen loss for ZoMaxx (0.29 ± 0.47mm) and Taxus (0.22 ± 0.41mm, p=NS) were similar, thus satisfying the primary angiographic endpoint. Secondary endpoints of the rates of in-segment and in-stent binary restenosis were also similar (5.9% vs. 5.8%, 7.8% vs. 7.9%, respectively).
Conclusions: At 9months, the ZoMaxx stent failed to achieve the primary endpoint of non-inferiority in TVR to the Taxus stent, but safety endpoints were equal between the two stent systems.
Trial registration: ClinicalTrials.gov NCT00140101.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.