Comparison of 3T and 7T MRI clinical sequences for ankle imaging

Eur J Radiol. 2012 Aug;81(8):1846-50. doi: 10.1016/j.ejrad.2011.05.023. Epub 2011 Jun 12.

Abstract

The purpose of this study was to compare 3T and 7T signal-to-noise and contrast-to noise ratios of clinical sequences for imaging of the ankles with optimized sequences and dedicated coils. Ten healthy volunteers were examined consecutively on both systems with three clinical sequences: (1) 3D gradient-echo, T(1)-weighted; (2) 2D fast spin-echo, PD-weighted; and (3) 2D spin-echo, T(1)-weighted. SNR was calculated for six regions: cartilage; bone; muscle; synovial fluid; Achilles tendon; and Kager's fat-pad. CNR was obtained for cartilage/bone, cartilage/fluid, cartilage/muscle, and muscle/fat-pad, and compared by a one-way ANOVA test for repeated measures. Mean SNR significantly increased at 7T compared to 3T for 3D GRE, and 2D TSE was 60.9% and 86.7%, respectively. In contrast, an average SNR decrease of almost 25% was observed in the 2D SE sequence. A CNR increase was observed in 2D TSE images, and in most 3D GRE images. There was a substantial benefit from ultra high-field MR imaging of ankles with routine clinical sequences at 7T compared to 3T. Higher SNR and CNR at ultra-high field MR scanners may be useful in clinical practice for ankle imaging. However, carefully optimized protocols and dedicated extremity coils are necessary to obtain optimal results.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Ankle / anatomy & histology*
  • Ankle Joint / anatomy & histology*
  • Female
  • Humans
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Reproducibility of Results
  • Sensitivity and Specificity