Improvement of freehand ultrasound calibration accuracy using the elevation beamwidth profile

Ultrasound Med Biol. 2011 Aug;37(8):1314-26. doi: 10.1016/j.ultrasmedbio.2011.05.004. Epub 2011 Jun 16.

Abstract

This article presents a novel approach that incorporates an ultrasound slice-thickness profile into a filtered, weighted-least-square framework to improve the reconstruction accuracy of a real-time freehand calibration system. An important part of the system is a slice-thickness calibration device that aids in the extraction of the slice thickness across a wide range of imaging depths. Extensive experiments were conducted on a 10,000-image dataset to evaluate the effects of the framework on the calibration accuracy. The results showed that three-dimensional (3-D) reconstruction errors were significantly reduced in every experiment (p < 0.001). Real-time testing showed that the proposed method worked effectively with a small number of input images, suggesting great potential for intraoperative use where only a limited number of data may be available. This new framework can enable efficient quality control of calibration accuracy in real-time operating-room use.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration*
  • Imaging, Three-Dimensional / instrumentation
  • Imaging, Three-Dimensional / standards*
  • Least-Squares Analysis
  • Operating Rooms
  • Phantoms, Imaging
  • Ultrasonography / instrumentation
  • Ultrasonography / standards*