Objectives: This study sought to evaluate the technical success and clinical outcomes of an adjunctive crossover balloon occlusion technique (CBOT) combined with the 10-F Prostar percutaneous closure device (PCD) on the incidence of vascular and bleeding complications in patients after transfemoral transcatheter aortic valve implantation (TAVI).
Background: Vascular closure following large-vessel access has most commonly been performed using a surgical cut-down and repair procedure.
Methods: Between November 2008 and September 2010, 58 consecutive patients with severe aortic stenosis underwent TAVI via a retrograde femoral artery approach using the Edwards-SAPIEN transcatheter valve. Among these patients, 56 were treated with a CBOT using the "pre-close" technique and the 10-F Prostar system. The technical success of this new CBOT and the 30-day frequency of clinical events, including all-cause mortality, major vascular complications, and major bleeding (defined according to a modified version of the Valve Academic Research Consortium criteria), were assessed.
Results: Successful closure was obtained in all but 3 patients (94.6%). The 30-day frequencies of mortality, major vascular complications, and major bleeding were 7.1%, 14.3%, and 5.4% respectively. No deaths were directly related to access site complications. Fourteen patients (25%) received at least 1 transfusion during the index hospitalization, of which 8 (57.1%) were not related to vascular complications. The mean and median hospital lengths of stay were 7.8 and 6.0 days.
Conclusions: This new percutaneous adjunctive CBOT combined with the Prostar PCD resulted in controlled, safe, and successful percutaneous closure in most patients after TAVI.
Copyright © 2011 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.