Compensation of slice profile mismatch in combined spin- and gradient-echo echo-planar imaging pulse sequences

Magn Reson Med. 2012 Feb;67(2):378-88. doi: 10.1002/mrm.23012. Epub 2011 Aug 19.

Abstract

Combined acquisition of gradient-echo and spin-echo signals in MRI time series reveals additional information for perfusion-weighted imaging and functional MRI because of differences in the sensitivity of gradient-echo and spin-echo measurements to the properties of the underlying vascular architecture. The acquisition of multiple echo trains within one time frame facilitates the simultaneous estimation of the transversal relaxation parameters R2 and R2*. However, the simultaneous estimation of these parameters tends to be incorrect in the presence of slice profile mismatches between signal excitation and subsequent refocusing pulses. It is shown here that improvements in pulse design reduced R2 and R2* estimation errors. Further improvements were achieved by augmented parameter estimation through the introduction of an additional parameter δ to correct for discordances in slice profiles to facilitate more quantitative measurements. Moreover, the analysis of time-resolved acquisitions revealed that the temporal stability of R2 estimates could be increased with improved pulse design, counteracting low contrast-to-noise ratios in spin-echo-based perfusion and functional MRI.

Publication types

  • Comparative Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Artifacts
  • Brain / anatomy & histology
  • Brain / blood supply
  • Echo-Planar Imaging / methods*
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Angiography / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Software Design