Color Doppler artifact in anechoic regions

J Ultrasound Med. 1990 May;9(5):255-60. doi: 10.7863/jum.1990.9.5.255.

Abstract

For color Doppler imaging, several types of signal processing are employed in order to produce acceptable images of blood flow in blood vessels while suppressing color in moving solid tissue. The processing can produce an artifact in which color may arise from noise or from tissue motion and fill anechoic regions preferentially. This artifact may complicate the differentiation of areas with blood flow from anechoic regions without flow. By using four different color Doppler ultrasound units to image a tissue-equivalent phantom containing anechoic cylinders, artifactual color resulted when gain was raised sufficiently. This color was concentrated in anechoic regions of a gray-scale image that did not contain flow. In two instruments, this artifact was only observed when the transducer was vibrated, simulating tissue motion. In these instruments, the identification of low-frequency, high-amplitude Doppler signals is used to locate moving solid tissue and so suppress color in these regions. In the other two instruments, the presence of echoes within the image suppressed the assignment of color. With both types of processing, color may appear artifactually in echo-free regions without flow, such as fluid collections. Presence or absence of flow should be confirmed by Doppler spectral analysis. An understanding of the origin and appearance of artifactual color can prevent its occurrence from detracting from the usefulness of color Doppler imaging.

MeSH terms

  • Aorta, Abdominal
  • Aortic Aneurysm / diagnosis
  • Color
  • Femoral Artery
  • Hematoma / diagnosis
  • Humans
  • Image Processing, Computer-Assisted*
  • Models, Structural
  • Regional Blood Flow*
  • Ultrasonography / methods*