High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: initial experience

Eur Radiol. 2012 Jan;22(1):129-37. doi: 10.1007/s00330-011-2257-5. Epub 2011 Sep 14.

Abstract

Objective: To investigate the feasibility, image quality and radiation dose for high-pitch dual-source CT angiography (CTA) of the whole aorta without ECG synchronisation.

Methods: Each group of 40 patients underwent CTA either on a 16-slice (group 1) or dual-source CT device with conventional single-source (group 2) or high-pitch mode with a pitch of 3.0 (group 3). The presence of motion or stair-step artefacts of the thoracic aorta was independently assessed by two readers.

Results: Subjective and objective scoring of motion and artefacts were significantly reduced in the high-pitch examination protocol (p < 0.05). The imaging length was not significantly different, but the imaging time was significantly (p < 0.001) shorter in the high-pitch group (12.2 vs. 7.4 vs. 1.7 s for groups 1, 2 and 3). The ascending aorta and the coronary ostia were reliably evaluable in all patients of group 3 without motion artefacts as well.

Conclusion: High-pitch dual-source CT angiography of the whole aorta is feasible in unselected patients. As a significant advantage over regular pitch protocols, motion-free imaging of the aorta is possible without ECG synchronisation. Thus, this CT mode bears potential to become a standard CT protocol before trans-catheter aortic valve implantation (TAVI).

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Aorta, Thoracic / diagnostic imaging*
  • Aorta, Thoracic / physiopathology
  • Aortic Diseases / diagnostic imaging*
  • Aortic Diseases / physiopathology
  • Artifacts*
  • Coronary Angiography* / methods
  • Feasibility Studies
  • Female
  • Heart Rate
  • Humans
  • Male
  • Middle Aged
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted*
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed* / methods