Performance of dry electrode with bristle in recording EEG rhythms across brain state changes

Annu Int Conf IEEE Eng Med Biol Soc. 2011:2011:59-62. doi: 10.1109/IEMBS.2011.6089896.

Abstract

In this paper we evaluate the physiological performance of a silver-silver chloride dry electrode with bristle (B-Electrode) in recording EEG data. For this purpose, we compare the performance of the bristle electrode in recording EEG data with the standard wet gold-plated cup electrode (G-Electrode) using two different brain state change tasks including resting condition with eyes-closed and performing mathematical task with eyes-open. Using a 2 channel recording device, eyes-closed command data were collected from each of 6 participants for a period of 20 sec and the same procedure was applied for the mathematical calculation task. These data were used for statistical and classification analyse. Although, B-electrode has shown a slightly higher performance compared with G-electrode in both tasks, but analyse did not reveal any significant differences between both electrodes in all six subjects tested.

MeSH terms

  • Adult
  • Biological Clocks / physiology*
  • Brain / physiology*
  • Electrodes*
  • Electroencephalography / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis