High-resolution alpha-particle spectrometry was performed on the (230)U decay series. A (230)U source was prepared on a stainless steel disc by electrodeposition in an ammonium nitrate solution. Spectrometry of the alpha-particle energy spectrum was performed with ion-implanted planar silicon detectors in vacuum. A set of alpha emission probabilities is presented for (230)U and (226)Th. The measured peak intensities were corrected mathematically for coincidental detection of alpha-particles and conversion electrons emitted in the same decay. A good agreement with literature data was observed. The uncertainty budget and the correlation matrix are presented. The validity of the alpha-particle energies was tested and could be confirmed for most peaks within a few keV, but discrepancies were found for the 2nd peak of (226)Th and the main peak of (218)Rn.
Copyright © 2012 Elsevier Ltd. All rights reserved.