Robustness of two MALDI-TOF mass spectrometry systems for bacterial identification

J Microbiol Methods. 2012 May;89(2):133-6. doi: 10.1016/j.mimet.2012.03.003. Epub 2012 Mar 8.

Abstract

MALDI-TOF-MS systems (Microflex-Bruker Daltonics/BioTyper™ and Axima-Assurance-Shimadzu/SARAMIS-AnagnosTec) were assessed for bacterial identification. Focusing on bacteria difficult to identify routinely, 296 strains were identified by molecular biology techniques as gold standard. MALDI-TOF-MS identification provided correct results at genus and species level for 94.9%, 83.4% and 83.8%, 65.9% with Biotyper and Saramis respectively.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Bacteria / chemistry*
  • Bacteria / classification*
  • Bacteriological Techniques / methods*
  • Sensitivity and Specificity
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*