Background: Plasma and urinary levels of D-lactate have been linked to the presence of diabetes. Previously developed techniques have shown several limitations to further evaluate D-lactate as a biomarker for this condition.
Methods: D- and L-lactate were quantified using ultraperformance liquid chromatography tandem mass spectrometry with labelled internal standard. Samples were derivatized with diacetyl-L-tartaric anhydride and separated on a C(18)-reversed phase column. D- and L-lactate were analysed in plasma and urine of controls, patients with inflammatory bowel disease (IBD), and patients with type 2 diabetes (T2DM).
Results: Quantitative analysis of D- and L-lactate was achieved successfully. Calibration curves were linear (r(2) > 0.99) over the physiological and pathophysiological ranges. Recoveries for urine and plasma were between 96% and 113%. Inter- and intra-assay variations were between 2% and 9%. The limits of detection of D-lactate and L-lactate in plasma were 0.7 μmol/L and 0.2 μmol/L, respectively. The limits of detection of D-lactate and L-lactate in urine were 8.1 nmol/mmol creatinine and 4.4 nmol/mmol creatinine, respectively. Plasma and urinary levels of D- and L-lactate were increased in patients with IBD and T2DM as compared with controls.
Conclusion: The presented method proved to be suitable for the quantification of D- and L-lactate and opens the possibility to explore the use of D-lactate as a biomarker.