Background: Near infrared spectroscopy (NIRS) has been suggested as a possible means for detecting perfusion deficits in patients with acute compartment syndrome (ACS).
Study objectives: To longitudinally examine NIRS in an ACS model to determine its responsiveness to decreasing perfusion pressure.
Methods: A NIRS sensor pad was placed under a tourniquet over the anterior compartment in the mid-tibia region on 20 volunteers. Initial perfusion pressures and NIRS values were recorded. The tourniquet pressure was sequentially raised by 10 mm Hg in 10-min intervals until systolic pressure was surpassed. NIRS values and perfusion pressure were determined at the end of each 10-min interval.
Results: There was no change in mean NIRS values from the initial baseline until 30 mm Hg of perfusion pressure was reached. Additionally, a statistically significant drop in mean NIRS values was observed as perfusion pressures dropped from 10 mm Hg to 0 mm Hg, and again with subsequent decreases of 10 mm Hg perfusion pressure until systolic pressure was surpassed.
Conclusions: These results coincide with previously published studies using alternative methods of measuring blood flow or perfusion. NIRS values were responsive to decreasing perfusion pressures over a longitudinal period of time in an ACS model. These results suggest that NIRS may be useful for continuous, non-invasive monitoring of patients for whom ACS is a concern. Additional studies on traumatized patients are required.