A heuristic statistical stopping rule for iterative reconstruction in emission tomography

Ann Nucl Med. 2013 Jan;27(1):84-95. doi: 10.1007/s12149-012-0657-5. Epub 2012 Oct 4.

Abstract

Objective: We propose a statistical stopping criterion for iterative reconstruction in emission tomography based on a heuristic statistical description of the reconstruction process.

Methods: The method was assessed for MLEM reconstruction. Based on Monte-Carlo numerical simulations and using a perfectly modeled system matrix, our method was compared with classical iterative reconstruction followed by low-pass filtering in terms of Euclidian distance to the exact object, noise, and resolution. The stopping criterion was then evaluated with realistic PET data of a Hoffman brain phantom produced using the GATE platform for different count levels.

Results: The numerical experiments showed that compared with the classical method, our technique yielded significant improvement of the noise-resolution tradeoff for a wide range of counting statistics compatible with routine clinical settings. When working with realistic data, the stopping rule allowed a qualitatively and quantitatively efficient determination of the optimal image.

Conclusions: Our method appears to give a reliable estimation of the optimal stopping point for iterative reconstruction. It should thus be of practical interest as it produces images with similar or better quality than classical post-filtered iterative reconstruction with a mastered computation time.

MeSH terms

  • Brain / diagnostic imaging
  • Image Processing, Computer-Assisted / methods*
  • Monte Carlo Method
  • Phantoms, Imaging
  • Positron-Emission Tomography / methods*
  • Signal-To-Noise Ratio
  • Statistics as Topic / methods*