A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

Med Phys. 2012 Nov;39(11):6757-63. doi: 10.1118/1.4758067.

Abstract

Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC).

Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system.

Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 ± 0.007 cGy∕mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% ± 3.8% and -0.1% ± 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%.

Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brachytherapy / methods*
  • Calibration
  • Carcinoma
  • Dose Fractionation, Radiation
  • Humans
  • Nasopharyngeal Carcinoma
  • Nasopharyngeal Neoplasms / radiotherapy*
  • Radiation Dosage*
  • Radiometry / methods*
  • Reproducibility of Results
  • Time Factors