Objectives: The objective of this analysis was to evaluate the safety and effectiveness of XIENCE V in acute myocardial infarction (AMI).
Background: The XIENCE V(®) Everolimus-eluting coronary stent was superior to the TAXUS(®) paclitaxel-eluting stent in angiographic and clinical outcomes in the SPIRIT II, III, and IV randomized controlled trials, but patients with AMI were excluded.
Methods: XIENCE V USA is a large, prospective, multicenter, real-world single-arm postmarket surveillance trial. Consecutive patients undergoing PCI with XIENCE V were enrolled. For this analysis, clinical outcomes in 673 patients presenting with AMI (STEMI, n = 125) were as compared to patients without AMI (n = 3528) at 1 year.
Results: At 1 year, ARC-defined stent thrombosis (ST) rates were 1.08% in AMI vs. 0.85% in the non-AMI group (P = 0.4987). The late ST (30 days-1 year) rates were 0.31% vs. 0.47% (AMI vs. non-AMI, P = 0.7551). Rates of target lesion revascularization (TLR) were 4.1% vs. 4.6% (P = 0.6104), and rates of target lesion failure (TLF) were 9.1% vs. 8.5%, (P = 0.5964). With the historical WHO definition of MI, 1 year TLF rates were 7.0% vs. 6.7% (P = 0.8001). Improvements in quality of life, angina frequency, angina stability, and physical limitations occurred at 6 months (each P < 0.0001) and were sustained at 1 year in both groups. There were no significant differences in clinical outcomes between STEMI and non-STEMI patients.
Conclusions: At 1 year, AMI patients treated with XIENCE V had low rates of ST, TLR, and TLF, similar to non-AMI patients. Marked improvements in patients' health status in this subgroup were also demonstrated.
Trial registration: ClinicalTrials.gov NCT00676520.
Keywords: drug eluting stent; everolimus; major adverse cardiac events; stent.
Copyright © 2012 Wiley Periodicals, Inc.