MRI-guided core biopsy of the prostate in the supine position--introduction of a simplified technique using large-bore magnet systems

Eur Radiol. 2013 May;23(5):1415-9. doi: 10.1007/s00330-012-2698-5. Epub 2012 Nov 24.

Abstract

Objectives: To introduce a simplified technique for MRI-guided core biopsies (MRGB) of the prostate in the supine position using large-bore magnet systems.

Methods: Fifty men with a history of negative transrectal ultrasound-guided biopsies underwent MRGB in either a 1.5-T (13/50) or 3.0-T (37/50) wide-bore MRI unit. MRGBs were conducted with the patients in a supine position using a dedicated MR-compatible biopsy device.

Results: We developed a dedicated positioning device for the supine position. Using this device, the biopsies were performed successfully in all patients. Apart from minor rectal bleeding, only one patient developed a major side effect (urosepsis). Histology revealed prostate cancer in 25/50 (50 %) patients.

Conclusions: The new technique appears feasible. Its major advantage is the more comfortable and patient-friendly supine position during the biopsy without the need to modify the MRI system's patient table.

Key points: • A novel positioning device for MRI-guided prostate biopsies has been developed. • Biopsies can be performed in the patient-friendly supine position. • The positioning device can be utilised without modifying the MRI's patient table.

MeSH terms

  • Aged
  • Equipment Design
  • Equipment Failure Analysis
  • Humans
  • Image Enhancement / instrumentation
  • Image Enhancement / methods
  • Image-Guided Biopsy / instrumentation
  • Image-Guided Biopsy / methods*
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Patient Positioning / methods*
  • Prostate / pathology*
  • Prostatic Neoplasms / pathology*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Supine Position