Background: The cobas TaqScreen MPX Test, version 2.0, a multiplex, multi-dye nucleic acid amplification technology (NAT) test from Roche was evaluated by two European Blood Banks, the German Red Cross Blood Donor Service, Frankfurt, Germany and Centro de Hemoterapia y Hemodonación de Castilla y León, Valladolid, Spain. In addition, the cobas TaqScreen DPX Test was evaluated for the simultaneous detection and quantitation of parvovirus B19 and the detection of hepatitis A virus (HAV).
Study design and methods: The performances of the two tests were evaluated regarding the analytical sensitivity, the reproducibility of the tests using samples containing low concentrations of each virus and cross-contamination using samples containing high titres of virus.
Results: The analytical sensitivity of the MPX Test, version 2.0, obtained by the German Red Cross Blood Donor Service was 1·1, 3·9 and 43·3 IU/ml for HBV, HCV and HIV-1, respectively. The comparable analytical sensitivity at Centro de Hemoterapia y Hemodonación de Castilla y León was 3·5, 17·6 and 50·6 IU/ml for HBV, HCV and HIV-1, respectively. The analytical sensitivity of the DPX test determined by the German Red Cross Blood Donor Service was 0·6 and 3·8 IU/ml for HAV and B19.
Conclusion: These multiplex and multi-dye blood screening assays represent a flexible NAT screening system for mini-pools between 6 and 96 samples per pool and fulfil all requirements of the European Pharmacopoeia for HCV and B19V testing of plasma for fractionation. The inclusion of a new multi-dye technology means discriminatory assays are no longer required for either test thus improving workflow, turn-around time and minimize the risk of obtaining a reactive result for which the virus cannot be identified.
© 2012 The Author(s). Vox Sanguinis © 2012 International Society of Blood Transfusion.