ECG-based signal analysis technologies for evaluating patients with acute coronary syndrome: a systematic review

J Electrocardiol. 2013 Mar-Apr;46(2):92-7. doi: 10.1016/j.jelectrocard.2012.11.010. Epub 2012 Dec 27.

Abstract

Background/objectives: Timely identification of cardiac ischemia is critical in patients with acute coronary syndrome (ACS). The first test is often the standard, resting 12-lead ECG. Given its limitations, signal analysis enhancements have been proposed. We summarize the published evidence for commercially available ECG-based signal analysis technologies.

Methods: This is a systematic review of the English-language published literature.

Results: Published evidence meeting inclusion criteria was available for two devices: PRIME ECG and LP 3000. Meta-analysis of eight studies estimated a 68.4% sensitivity (95% CI, 35.1%-89.7%) and 91.4% specificity (CI, 83.6%-95.7%) for the PRIME ECG, compared with 40.5% sensitivity (CI, 19.6%-65.5%) and 95.0% specificity (CI, 87.9%-98.0%) for the standard 12-lead ECG.

Conclusions: Existing evidence is insufficient to confidently inform the appropriate use of ECG-based signal analysis technologies for detecting ischemia or infarct in ACS. Further research is needed to determine in what circumstances, if any, these devices might precede, replace, or add to the standard ECG in test strategies for detecting ischemia or infarct in ACS.

Publication types

  • Comparative Study
  • Meta-Analysis
  • Research Support, U.S. Gov't, P.H.S.
  • Review
  • Systematic Review

MeSH terms

  • Acute Coronary Syndrome / diagnosis*
  • Acute Coronary Syndrome / epidemiology*
  • Diagnosis, Computer-Assisted / methods*
  • Diagnosis, Computer-Assisted / statistics & numerical data*
  • Electrocardiography / methods*
  • Electrocardiography / statistics & numerical data*
  • Humans
  • Prevalence
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted