Development of a sensitive and selective LC-MS/MS method for the determination of urea in human epithelial lining fluid

J Chromatogr B Analyt Technol Biomed Life Sci. 2013 Feb 15:917-918:24-9. doi: 10.1016/j.jchromb.2012.11.035. Epub 2013 Jan 9.

Abstract

A sensitive, selective, and quantitative method for the determination of urea has been developed and validated in human epithelial lining fluid (ELF; the supernatant from bronchoalveolar lavage). The method employs a simple derivatization of urea with camphanic chloride to improve the chromatographic retention and separation. The derivatization was performed after drying an aliquot of ELF (20μL) without prior sample clean-up. Ultra High Performance Liquid Chromatography (UHPLC) on a HSS-T3 stationary phase column with 1.8μm particle size was used for chromatographic separation coupled to tandem mass spectrometry. The method was validated over the concentration range of 8.78-103.78μg/mL, however the dynamic range can be further lowered if needed. The results from assay validation show that the method is rugged, precise, accurate, and well-suited to support analysis of urea in ELF samples. In addition, the relatively small sample volume (20μL) and a run time of 1.5min facilitate automation and allow for high-throughput analysis. This derivatization method was compared to a commercially available colorimetric assay kit, and it was used in a preclinical non-GLP mouse study where urea measurements were used as marker of bronchoalveolar lavage fluid dilution.

MeSH terms

  • Animals
  • Bronchoalveolar Lavage Fluid / chemistry*
  • Chromatography, High Pressure Liquid / methods*
  • Colorimetry
  • Humans
  • Linear Models
  • Mice
  • Sensitivity and Specificity
  • Tandem Mass Spectrometry / methods*
  • Urea / analysis*
  • Urea / chemistry

Substances

  • Urea