High-resolution peptide separations using nano-LC at ultra-high pressure

J Sep Sci. 2013 Apr;36(7):1192-9. doi: 10.1002/jssc.201201087. Epub 2013 Mar 4.

Abstract

We report on the optimization of nano-LC gradient separations of proteomic samples that vary in complexity. The gradient performance limits were visualized by kinetic plots depicting the gradient time needed to achieve a certain peak capacity, while using the maximum system pressure of 80 MPa. The selection of the optimal particle size/column length combination and corresponding gradient steepness was based on scouting the performance of 75 μm id capillary columns packed with 2, 3, and 5 μm fully porous silica C18 particles. At optimal gradient conditions, peak capacities up to 500 can be obtained within a 120 min gradient using 2 μm particle-packed capillary columns. Separations of proteomic samples including a cytochrome c tryptic digest, a bovine serum albumin tryptic digest, a six protein mix digest, and an Escherichia coli digest are demonstrated while operating at the kinetic-performance limit, i.e. using 2-μm packed columns, adjusting the column length and scaling the gradient steepness according to sample complexity. Finally, good run-to-run retention time stability (RSD values below 0.18%) was demonstrated applying ultra-high pressure conditions.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid*
  • Nanotechnology / methods*
  • Particle Size
  • Peptides / chemistry*
  • Peptides / isolation & purification

Substances

  • Peptides