Purpose: This study aims to report on the measured in vivo contribution of soft lenses of various powers to the optics of a piggyback system (PBS).
Methods: This prospective, non-dispensing clinical study was conducted on regular wearers of contact lenses who showed regular corneal profiles. Subjects were masked to the products used. The study involved the use of a spherical soft lens of three different powers in a PBS, used as a carrier for a rigid gas permeable lens. Baseline data were collected and soft lenses were then fitted on both eyes of each subject. Both lenses were assessed for position and movement. Over-refraction was obtained. Soft lens power contribution to the optics (SLPC) of a PBS system was estimated by computing initial ametropia, lacrymal lens, rigid lens powers and over-refraction. A set of data on one eye was kept, for each subject, for statistical analysis.
Results: Thirty subjects (12 males, 18 females), aged 24.4 (±4.5) years, were enrolled. The use of plus powered soft lenses enhanced initial RGP lens centration. Once optimal fit was achieved, all lenses showed normal movement. SLPC represented 21.3% of the initial soft lens power when using a -6.00 carrier, and 20.6% for a +6.00. A +0.50 did not contribute to any power induced in the system. These results are generally in accordance with theoretical model developed in the past.
Conclusion: On average, except for the low-powered carrier, the use of a spherical soft lens provided 20.9% of its marked power. To achieve better results, the use of a plus-powered carrier is recommended.
Keywords: Astigmatism; Contact lens optics; Piggyback systems; Rigid gas-permeable lenses.
Copyright © 2013 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.