Monte Carlo optimization of total body irradiation in a phantom and patient geometry

Phys Med Biol. 2013 Apr 21;58(8):2461-9. doi: 10.1088/0031-9155/58/8/2461. Epub 2013 Mar 21.

Abstract

The objective of this work is to apply a Monte Carlo (MC) accelerator model, validated by experimental data at isocentre distances, to a large-field total body irradiation (TBI) technique and to develop a strategy for individual patient treatment on the basis of MC dose distributions. Calculations are carried out using BEAMnrc/DOSXYZnrc code packages for a 15 MV Varian accelerator. Acceptable agreement is obtained between MC data and measurements in a large water phantom behind a spoiler at source-skin distances (SSD) = 460 cm as well as in a CIRS® thorax phantom. Dose distributions in patients are studied when simulating bilateral beam delivery at a distance of 480 cm to the patient central sagittal plane. A procedure for individual improvement of the dose uniformity is suggested including the design of compensators in a conventional treatment planning system (TPS) and a subsequent update of the dose distribution. It is demonstrated that the dose uniformity for the simple TBI technique can be considerably improved. The optimization strategy developed is straightforward and suitable for clinics where the TPS available is deficient to calculate 3D dose distributions at extended SSD.

MeSH terms

  • Acceleration
  • Humans
  • Models, Theoretical
  • Monte Carlo Method*
  • Phantoms, Imaging*
  • Radiotherapy Dosage
  • Thorax / radiation effects
  • Whole-Body Irradiation / instrumentation*