Rationale: Neurological dysfunction is common in primary coenzyme Q10 (2,3-dimethoxy, 5-methyl, 6-polyisoprene parabenzoquinone; CoQ10 ; ubiquinone) deficiencies, the most readily treatable subgroup of mitochondrial disorders. Therapeutic benefit from CoQ10 supplementation has also been noted in other neurodegenerative diseases. CoQ10 can be measured by high-performance liquid chromatography (HPLC) in plasma, muscle or leucocytes; however, there is no reliable method to quantify CoQ10 in cerebrospinal fluid (CSF). Additionally, many methods use CoQ9 , an endogenous ubiquinone in humans, as an internal standard.
Methods: Deuterated CoQ10 (d6 -CoQ10 ) was synthesised by a novel, simple, method. Total CoQ10 was measured by liquid chromatography/tandem mass spectrometry (LC/MS/MS) using d6 -CoQ10 as internal standard and 5 mM methylamine as an ion-pairing reagent. Chromatography was performed using a Hypsersil GOLD C4 column (150 × 3 mm, 3 µm).
Results: CoQ10 levels were linear over a concentration range of 0-200 nM (R(2) = 0.9995). The lower limit of detection was 2 nM. The inter-assay coefficient of variation (CV) was 3.6% (10 nM) and 4.3% (20 nM), and intra-assay CV 3.4% (10 nM) and 3.6% (20 nM). Reference ranges were established for CoQ10 in CSF (5.7-8.7 nM; n = 17), fibroblasts (57.0-121.6 pmol/mg; n = 50) and muscle (187.3-430.1 pmol/mg; n = 15).
Conclusions: Use of d6 -CoQ10 internal standard has enabled the development of a sensitive LC/MS/MS method to accurately determine total CoQ10 levels. Clinical applications of CSF CoQ10 determination include identification of patients with cerebral CoQ10 deficiency, and monitoring CSF CoQ10 levels following supplementation.
Copyright © 2013 John Wiley & Sons, Ltd.