Multicenter R2* mapping in the healthy brain

Magn Reson Med. 2014 Mar;71(3):1103-7. doi: 10.1002/mrm.24772.

Abstract

Purpose: The R2* relaxation rate constant has been suggested as a sensitive measure for iron accumulation. The aim of this multi-center study was to assess the inter-scanner and inter-subject variability of R2* mapping and to investigate the relationship between brain volume and R2* in specific structures.

Methods: R2* mapping was performed in 81 healthy subjects in seven centers using different 3 T systems. R2* was calculated from a dual-echo gradient echo sequence and was assessed in several deep gray matter structures. The inter-scanner and inter-subject variability of R2* was calculated by means of the coefficient of variation before and after correcting for age.

Results: Significant center effects were seen in some regions which get lost after age correction. The coefficient of variation for the inter-center variability was much lower (<5.6%) than for the intra-subject variability (6.7%-11.7%). R2* in the putamen and red nucleus scaled with cortical volume while R2* in the globus pallidus and the substantia nigra was negatively associated with white matter volume.

Conclusion: R2* is a robust and reproducible measure in a multicenter setting provided that a standardized MRI protocol is used. The relationship between iron concentration in deep gray matter and volume of specific brain compartments needs further investigation.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Brain / anatomy & histology*
  • Europe
  • Female
  • Gray Matter / anatomy & histology*
  • Humans
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Organ Size
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • White Matter / anatomy & histology*
  • Young Adult