Dependence of blood T(2) on oxygenation at 7 T: in vitro calibration and in vivo application

Magn Reson Med. 2014 Jun;71(6):2035-42. doi: 10.1002/mrm.24868. Epub 2013 Jul 10.

Abstract

Purpose: The calibratable relationship between blood oxygenation (Y) and T(2) allows quantification of cerebral venous oxygenation. We aim to establish a calibration plot between blood T(2) , Y, and hematocrit at 7 T, and using T(2) -relaxation-under-spin-tagging MRI, determine human venous blood oxygenation in vivo.

Methods: In vitro experiments were performed at 7 T on bovine blood samples using a Carr-Purcell-Meiboom-Gill-T2 sequence, from which we characterized the relationship among T(2) , Y, and hematocrit. T(2) -relaxation-under-spin-tagging MRI was implemented at 7 T to measure venous blood T2 in vivo, from which oxygenation was estimated using the in vitro calibration plot. Hyperoxia was performed to test the sensitivity of the method to oxygenation changes, and the 7 T results were compared with those at 3 T.

Results: In vitro data showed that arterial and venous T(2) at 7 T are 68 and 20 ms, respectively, at a typical hematocrit of 0.42. In vivo measurement showed a cerebral venous oxygenation of 64.7 ± 5.0% and a test-retest coefficient-of-variation of 3.6 ± 2.4%. Hyperoxia increased Yv by 9.0 ± 1.4% (P = 0.001) and the 3 and 7 T results showed a strong correlation (R = 0.95) across individuals.

Conclusion: We provided an in vitro calibration plot for conversion of blood T(2) to oxygenation at 7 T and demonstrated its utility in vivo.

Keywords: 7 T; TRUST; blood T2; blood oxygen saturation; brain.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Blood Flow Velocity / physiology
  • Brain / blood supply*
  • Brain / physiology
  • Calibration
  • Cattle
  • Cerebrovascular Circulation / physiology*
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods*
  • In Vitro Techniques
  • Magnetic Resonance Imaging / methods*
  • Oxygen / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Oxygen