GESFIDE-PROPELLER approach for simultaneous R2 and R2* measurements in the abdomen

Magn Reson Imaging. 2013 Dec;31(10):1760-5. doi: 10.1016/j.mri.2013.08.003. Epub 2013 Sep 14.

Abstract

Purpose: To investigate the feasibility of combining GESFIDE with PROPELLER sampling approaches for simultaneous abdominal R2 and R2* mapping.

Materials and methods: R2 and R2* measurements were performed in 9 healthy volunteers and phantoms using the GESFIDE-PROPELLER and the conventional Cartesian-sampling GESFIDE approaches.

Results: Images acquired with the GESFIDE-PROPELLER sequence effectively mitigated the respiratory motion artifacts, which were clearly evident in the images acquired using the conventional GESFIDE approach. There was no significant difference between GESFIDE-PROPELLER and reference MGRE R2* measurements (p=0.162) whereas the Cartesian-sampling based GESFIDE methods significantly overestimated R2* values compared to MGRE measurements (p<0.001).

Conclusion: The GESFIDE-PROPELLER sequence provided high quality images and accurate abdominal R2 and R2* maps while avoiding the motion artifacts common to the conventional Cartesian-sampling GESFIDE approaches.

Keywords: GESFIDE; PROPELLER; R2/R2* mapping; abdominal.