Objective: We performed a multicenter study to calculate the upper reference limits (URL) for urine particle quantification in mid-stream samples by using automated urine analyzers.
Design & methods: Two laboratories tested 283 subjects using a Sysmex UF-100, two other laboratories tested 313 subjects using Sysmex UF-1000i, whereas two other laboratories tested 267 subjects using Iris IQ®200.
Results: The URLs of UF-100 in females and males were 7.8/μL and 6.7/μL for epithelial cells (EC), 11.1/μL and 9.9/μL for red blood cells (RBC), 10.2/μL and 9.7/μL for white blood cells (WBC), and 0.85/μL and 0.87/μL for cylinders (CAST). The URLs of UF-1000i in females and males were 7.6/μL and 7.1/μL for EC, 12.2/μL and 11.1/μL for RBC, 11.9/μL and 11.7/μL for WBC, and 0.88/μL and 0.86/μL for CAST. The URLs of Iris IQ®200 in females and males were 7.8/μL and 6.6/μL for EC, 12.4/μL and 10.1/μL for RBC, 10.9/μL and 9.9/μL for WBC, and 1.1/μL and 1.0/μL for CAST.
Conclusion: The URLs obtained in this study were comparable to the lowest values previously reported in the literature. Moreover, no gender-related difference was observed, and analyzer-specific upper reference limits were very similar.
Keywords: Automation; CAST; CASTS/cylinders; CI 90%; Confidence Index 90%; EC; IQR; Inter Quartile Range; Multicentre study; RBC; Red Blood Cells/erythrocytes; Reference values; Squamous Epithelial Cells; URL; Upper Reference Limits; Urine particles; WBC; White Blood Cells/leukocytes.