Quantification of linezolid in serum by LC-MS/MS using semi-automated sample preparation and isotope dilution internal standardization

Clin Chem Lab Med. 2014 Mar;52(3):381-9. doi: 10.1515/cclm-2013-0594.

Abstract

Background: Linezolid serum concentrations have been shown to be highly variable in critically ill patients with often sub-therapeutic drug levels regarding minimal inhibitory concentrations for relevant pathogens. Consequently, therapeutic drug monitoring of linezolid must be considered, requiring a reliable and convenient analytical method. We therefore developed and validated an LC-MS/MS method applying isotope dilution internal standardization and on-line solid phase extraction for serum linezolid quantification.

Methods: Sample preparation was based on protein precipitation and on-line solid phase extraction with two-dimensional liquid chromatography and column switching. Three-fold deuterated linezolid was used as the internal standard. The method was validated involving two separate LC-MS/MS systems covering the concentration range of 0.13-32 mg/L. The run time was 4 min.

Results: Validation revealed good analytical performance, with inaccuracy <6% and imprecision of <7.3% (CV) for six quality control samples (0.38-16.0 mg/L). The method was found to be robust during the validation process and during a pharmacokinetic study so far involving 600 samples. Comparative measurements on two LC-MS/MS systems revealed close agreement.

Conclusions: This LC-MS/MS assay described herein is a convenient, robust and reliable method for linezolid quantification in serum which can be routinely applied using different LC-MS/MS systems. The method can be used for clinical studies and subsequent TDM of linezolid.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acetamides / blood*
  • Acetamides / isolation & purification*
  • Analytic Sample Preparation Methods / methods*
  • Automation
  • Blood Chemical Analysis / methods*
  • Blood Chemical Analysis / standards*
  • Chromatography, Liquid
  • Humans
  • Linezolid
  • Oxazolidinones / blood*
  • Oxazolidinones / isolation & purification*
  • Radioisotope Dilution Technique
  • Reference Standards
  • Reproducibility of Results
  • Solid Phase Extraction
  • Tandem Mass Spectrometry*

Substances

  • Acetamides
  • Oxazolidinones
  • Linezolid