Installation and performance testing of an XtRa-NaI(Tl) Compton Suppression System at the NED-NTUA

Appl Radiat Isot. 2014 May:87:361-4. doi: 10.1016/j.apradiso.2013.11.056. Epub 2013 Nov 27.

Abstract

This paper presents the Compton Suppression System, recently installed at the Nuclear Engineering Department of NTUA. The system consists of an XtRa Ge detector coupled with a NaI(Tl) guard detector. The electronic set-up allows for the simultaneous collection of both the suppressed and the unsuppressed spectra. System performance is investigated using certified point and volume sources. Parameters such as Peak Suppression Factors, peak-to-Compton ratios and minimum detectable activity for specific radionuclides are determined.

Keywords: Compton suppression system; Gamma-ray spectrometry; Performance testing.