The recently launched Liaison XL Murex HIV Ab/Ag assay (DiaSorin S.p.A) uses chemiluminescence immunoassay technology for the combined qualitative determination of p24 antigen of HIV-1 and specific antibodies to both HIV-1 and HIV-2. We studied 571 serum samples from those submitted to our laboratory for HIV screening. The samples were divided into 3 subsets: subset A, 365 samples collected prospectively during 1 week; subset B, 158 samples from confirmed HIV-positive patients; and subset C, 48 samples with a positive screening result but a negative or indeterminate confirmatory test result. Our standard screening/confirmatory algorithm was used as a reference. In subset A (prospective), 5 samples were positive and 360 negative by the standard procedure. Liaison XL Murex HIV Ab/Ag correctly identified all 5 positive samples (100%) and 357 negative samples (99.2%). In subset B (confirmed positive), all 158 positive samples were in total agreement in both procedures. In subset C (screen positive only), Liaison XL Murex HIV Ab/Ag yielded accurate results in 42 out of 48 samples (87.5%). Global sensitivity and specificity for Liaison XL Murex HIV Ab/Ag (all subsets included) were 98.3% and 98.5%, respectively. Considering only nonselected prospective samples and confirmed positive samples (subsets A and B), the corresponding sensitivity and specificity values were 100% and 99.2%, respectively. The new fully automated HIV screening test showed high sensitivity and specificity compared to our standard algorithm. Its added advantage of being able to detect HIV-1 and HIV-2 antibodies and p24 antigen separately could prove useful in the diagnosis of early infections.