Dosimetric audit in brachytherapy

Br J Radiol. 2014 Sep;87(1041):20140105. doi: 10.1259/bjr.20140105. Epub 2014 May 6.

Abstract

Dosimetric audit is required for the improvement of patient safety in radiotherapy and to aid optimization of treatment. The reassurance that treatment is being delivered in line with accepted standards, that delivered doses are as prescribed and that quality improvement is enabled is as essential for brachytherapy as it is for the more commonly audited external beam radiotherapy. Dose measurement in brachytherapy is challenging owing to steep dose gradients and small scales, especially in the context of an audit. Several different approaches have been taken for audit measurement to date: thimble and well-type ionization chambers, thermoluminescent detectors, optically stimulated luminescence detectors, radiochromic film and alanine. In this work, we review all of the dosimetric brachytherapy audits that have been conducted in recent years, look at current audits in progress and propose required directions for brachytherapy dosimetric audit in the future. The concern over accurate source strength measurement may be essentially resolved with modern equipment and calibration methods, but brachytherapy is a rapidly developing field and dosimetric audit must keep pace.

Publication types

  • Review

MeSH terms

  • Brachytherapy / instrumentation
  • Brachytherapy / standards*
  • Calibration
  • Humans
  • Medical Audit
  • Radiometry / methods
  • Radiotherapy Dosage
  • Thermoluminescent Dosimetry