Measurement of unimolecular decay in peptides of masses greater than 1200 units by a reflecting time-of-flight mass spectrometer

Rapid Commun Mass Spectrom. 1989 Dec;3(12):443-8. doi: 10.1002/rcm.1290031210.

Abstract

Daughter ions from decomposition of [M + H]+ parent ions have been observed in a time-of-flight mass spectrometer fitted with an ion mirror. Unit mass resolution is obtained for parent ions of masses up to several thousand u when the mirror voltage is set at a value determined by the usual velocity-focusing criterion for parent ions. Under this condition the daughter-ion resolution in the low-mass range suffers. Considerable improvement is obtained when the daughter-ion spectrum is examined in several segments, with the mirror voltage optimized for each segment. The daughter-ion spectrum from decay of metastable [M + H]+ parent ions in Substance P is measured using this technique. Reasons are suggested for differences between the spectrum obtained and a spectrum for the same compound recently obtained using a tandem double-focusing instrument.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Endorphins / chemistry*
  • Enkephalin, Leucine / chemistry*
  • Mass Spectrometry / methods
  • Protein Precursors / chemistry*
  • Substance P / chemistry*

Substances

  • Endorphins
  • Protein Precursors
  • Substance P
  • Enkephalin, Leucine
  • alpha-neoendorphin