Background: Coronary plaque size modification, by either local (device) or systemic treatments, has been the target for many years.
Methods: From ABSORB Cohort A (Absorb BVS 1.0), ABSORB Cohort B (Absorb BVS 1.1), SPIRIT FIRST (Multi-Link Vision vs. Xience V) & SPIRIT II (Xience V vs. Taxus), we calculated the total plaque area (vessel minus lumen area - thus it comprises both compartments - the plaque behind struts and the neointima.) changes by IVUS.
Results: A total of 313 patients were included. Comparison-at-6-month follow-up: All devices induced an increase in the total plaque area. The largest increase occurred with Vision and Taxus stents as compared to other devices [Absorb BVS (1.0 and 1.1) and Xience V], (p=0.0002). Comparison-at-2-year follow-up: Absorb BVS 1.1 had a larger increase from post procedure in total plaque compared to Absorb BVS 1.0, Xience V and Taxus (p=0.0499). However, in Absorb BVS 1.1 total plaque showed a reduction of 2.2% from 1 to 3 years. Specifically, the total plaque in the sequential cohorts of Absorb BVS 1.1 increased 16.2% from baseline to 2 years (Cohort B1) while at 3 years this increase is only 5% compared to baseline (Cohort B2).
Conclusions: Local devices affect coronary plaque size differently and it depends on the platform (metallic vs. polymeric) and on whether it is a bare - or drug eluting stent. Coronary scaffolds appear to be a promising alternative to metallic stents since they allow plaque regression at long-term follow-up.
Keywords: Bioresorbable scaffolds; Coronary atherosclerosis; Stents.