[Analysis for 2 samples with HLA-DQB1 allele dropout at exon 2 in sequence-based typing]

Zhonghua Yi Xue Yi Chuan Xue Za Zhi. 2014 Aug;31(4):496-8. doi: 10.3760/cma.j.issn.1003-9406.2014.04.019.
[Article in Chinese]

Abstract

Objective: To explore the reason for HLA-DQB1 allele dropout during routine sequence-based typing(SBT) in order to improve the accuracy of typing.

Methods: Two thousand samples derived from HLA high-resolution typing laboratory were typed for HLA-DQB1 locus using an AlleleSEQR HLA-DQB1 SBT kit. Non-conclusive results and "abnormal" sequencing samples were retyped using a LABType rSSO HD HLA-DQB1 kit and further analyzed with both sequence-specific primers and group-specific primers and sequenced for haplotype analysis.

Results: Among the 2000 samples, 2 samples with no conclusive result were identified. The heterozygosity was confirmed with both the LAB Type SSO HD HLA-DQB1 kit and PCR-SBT in house method. Subsequent HLA-DQB1 cloning and haplotype sequencing have elucidated that HLA-DQB1*02:02 dropped out at exon 2 for the first sample and HLA-DQB1*02:01:01 dropped out at exon 2 for the second sample during PCR amplification. No novel nucleotide mutation was found.

Conclusion: Our results indicated that preferential amplification at exon 2 of DQB1 may result in allele dropout in exon 2 sequences during HLA-DQB1 SBT test. This may provide useful information for HLA genotyping.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Alleles
  • DNA Primers / genetics
  • Exons*
  • Genotype
  • HLA-DQ beta-Chains / genetics*
  • Histocompatibility Testing / methods*
  • Humans
  • Polymerase Chain Reaction / methods*

Substances

  • DNA Primers
  • HLA-DQ beta-Chains
  • HLA-DQB1 antigen