Long-term clinical effects of ventricular pacing reduction with a changeover mode to minimize ventricular pacing in a general pacemaker population

Eur Heart J. 2015 Jan 14;36(3):151-7. doi: 10.1093/eurheartj/ehu336. Epub 2014 Sep 1.

Abstract

Aim: Right ventricular pacing (VP) has been hypothesized to increase the risk in heart failure (HF) and atrial fibrillation (AF). The ANSWER study evaluated, whether an AAI-DDD changeover mode to minimize VP (SafeR) improves outcome compared with DDD in a general dual-chamber pacemaker population.

Methods and results: ANSWER was a randomized controlled multicentre trial assessing SafeR vs. standard DDD in sinus node disease (SND) or AV block (AVB) patients. After a 1-month run-in period, they were randomized (1 : 1) and followed for 3 years. Pre-specified co-primary end-points were VP and the composite of hospitalization for HF, AF, or cardioversion. Pre-specified secondary end-points were cardiac death or HF hospitalizations and cardiovascular hospitalizations. ANSWER enrolled 650 patients (52.0% SND, 48% AVB) at 43 European centres and randomized in SafeR (n = 314) or DDD (n = 318). The SafeR mode showed a significant decrease in VP compared with DDD (11.5 vs. 93.6%, P < 0.0001 at 3 years). Deaths and syncope did not differ between randomization arms. No significant difference between groups [HR = 0.78; 95% CI (0.48-1.25); P = 0.30] was found in the time to event of the co-primary composite of hospitalization for HF, AF, or cardioversion, nor in the individual components. SafeR showed a 51% risk reduction (RR) in experiencing cardiac death or HF hospitalization [HR = 0.49; 95% CI (0.27-0.90); P = 0.02] and 30% RR in experiencing cardiovascular hospitalizations [HR = 0.70; 95% CI (0.49-1.00); P = 0.05].

Conclusion: SafeR safely and significantly reduced VP in a general pacemaker population though had no effect on hospitalization for HF, AF, or cardioversion, when compared with DDD.

Keywords: Atrial fibrillation; Dual-chamber pacing; Heart failure; Minimization of ventricular pacing; Randomized controlled trial; SafeR.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Atrioventricular Block / mortality
  • Atrioventricular Block / therapy*
  • Cardiac Pacing, Artificial / adverse effects
  • Cardiac Pacing, Artificial / methods*
  • Cardiac Pacing, Artificial / mortality
  • Death, Sudden, Cardiac / etiology
  • Electric Countershock / mortality
  • Electric Countershock / statistics & numerical data
  • Equipment Failure
  • Female
  • Hospitalization / statistics & numerical data
  • Humans
  • Male
  • Pacemaker, Artificial / adverse effects
  • Prospective Studies
  • Sick Sinus Syndrome / mortality
  • Sick Sinus Syndrome / therapy*
  • Single-Blind Method
  • Treatment Outcome