Background: Capsule endoscopy has been proposed as an alternative to fibreoptic endoscopy for oesophageal varices evaluation in cirrhotics. However, it shows only moderate sensitivity compared to fibreoptic endoscopy.
Aim: To compare post-meal capsule endoscopy to fibreoptic endoscopy, based on the hypothesis that meal-induced increase of portal pressure can enhance its sensitivity.
Methods: Twenty-five patients were submitted to fibreoptic endoscopy and, after a standard meal, capsule endoscopy.
Results: Post-meal capsule endoscopy detected varices in the 18 patients in whom fibreoptic endoscopy detected varices plus 3 more subjects (sensitivity 100%, specificity 70%); large varices in the 4 patients in whom fibreoptic endoscopy graded varices as large, plus 5 more subjects; red markers in the 5 patients in whom fibreoptic endoscopy detected red markers, plus 3 more subjects. High-risk varices were identified in 11 patients by post-meal capsule endoscopy and in 10 by fibreoptic endoscopy (sensitivity 100%, specificity 93.8%).
Conclusions: Post-meal capsule endoscopy identified more varices, large varices and red markers than fibreoptic endoscopy. The two methods detected similar proportions of high-risk varices. These data suggest that a standard meal can enhance the sensitivity of capsule endoscopy in the detection and grading of oesophageal varices in cirrhotics.
Keywords: Capsule endoscopy; Oesophageal varices; Portal hypertension.
Copyright © 2014 Editrice Gastroenterologica Italiana S.r.l. Published by Elsevier Ltd. All rights reserved.