Transcatheter aortic valve-in-valve implantation: clinical outcome as defined by VARC-2 and postprocedural valve dysfunction according to the primary mode of bioprosthesis failure

J Invasive Cardiol. 2014 Oct;26(10):542-7.

Abstract

Objectives: The objectives of this study were to investigate: (1) the clinical outcome of transcatheter aortic valve-in-valve (VIV) implantation according to Valve Academic Research Consortium (VARC)-2 criteria; and (2) to determine whether postprocedural transvalvular gradients differ in patients with bioprosthesis regurgitation or stenosis as primary mode of failure.

Background: Transcatheter aortic VIV implantation has become a feasible option for selected high-risk patients with failed aortic surgical bioprostheses.

Methods: Transcatheter aortic VIV implantation was performed in 14 high-risk individuals at the University of Zurich and University College London.

Results: The prosthesis was successfully implanted in 13 patients (93%). In 1 patient, a second transcatheter valve needed to be implanted due to valve malpositioning. Thirty-day all-cause mortality was 7% (1/14). Prosthetic valve dysfunction according to VARC-2 at 30 days was observed in 7/14 patients (50%) due to an increased postprocedural transvalvular gradient >20 mm Hg. Preprocedural transaortic gradients correlated significantly with postprocedural gradients (r=0.91; P<.001). At 30-day follow-up, postprocedural gradients were higher in patients with aortic stenosis as primary mode of failure as compared to those with aortic regurgitation (36 ± 6 mm Hg vs 16 ± 4 mm Hg; P=.01). None of the patients exhibited prosthetic valve regurgitation of more than mild degree.

Conclusion: The feasibility and safety of VIV implantation in failed aortic bioprostheses is demonstrated. A higher postprocedural gradient was observed after VIV implantation in patients with aortic stenosis as compared to regurgitation as primary mode of failure.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Bioprosthesis*
  • Feasibility Studies
  • Female
  • Hospitals, University
  • Humans
  • London
  • Male
  • Middle Aged
  • Postoperative Complications / etiology*
  • Postoperative Complications / surgery*
  • Prosthesis Failure*
  • Reoperation / mortality
  • Research
  • Switzerland
  • Transcatheter Aortic Valve Replacement / methods*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome